Název * : Fatigue testing of medial plates for treatment of distal tibia fractures
Citace : Frydrýšek, K., Theisz, G., Demel, J., Pleva, L., Pompach, M. a Fojtík, F. Fatigue testing of medial plates for treatment of distal tibia fractures. In: 54th International Conference on Experimental Stress Analysis: 54th International Conference on Experimental Stress Analysis 2016-05-30 Hotel Srní. Kašperské Hory: Department of Mechanics, Faculty of Applied Sciences, University of West Bohemia, 2016. s. 1-6. ISBN 9788026106241.
Podnázev :
Rok * : 2016
Obor * : Chirurgie vč. transplantologie
Počet stran * : 6
Strana od * : 1
Strana do * : 6
Forma vydání * : Tištená verze
Kód ISBN * : 9788026106241
Kód ISSN :
Název sborníku * : 54th International Conference on Experimental Stress Analysis
Sborník : Mezinárodní
Název nakladatele * : Department of Mechanics, Faculty of Applied Sciences, University of West Bohemia
Místo vydání * : Kašperské Hory
Stát vydání : Sborník vydaný v ČR
Název konference : 54th International Conference on Experimental Stress Analysis
Místo konání konference * : Hotel Srní
Datum zahájení konference * :
Typ akce podle státní
příslušnosti účastníků akce * :
Evropská akce
Kód UT WoS :
EID :
Klíčová slova anglicky * :
Biomechanics; Bone; Fatigue testing; Fracture; Internal fixation; LCP; Medial plate; Tibia
Popis v původním jazyce * :
The focus of the article is the biomechanical analysis of new locking compression medial plates. These types of plates are used in orthopedics and traumatology for internal fixation of fractures of the distal tibia. Low-cycle fatigue analyses were performed by repeated axial loading for non-fused bone with complicated comminuted fractures in the distal metaphysis. This type of fracture represents an extreme situation, where all loads between two fragments of the broken bones are carried by the plate. It is also essential to perform fatigue tests before clinical application in hospitals. From the results of the tests, the durability, loadings and fatigue crack parameters were determined and a reliability assessment was performed.
Popis v anglickém jazyce * :
Typ zdroje financování výsledku * : Jiné veřejné zdroje
Seznam projektů :
ID Projektu Název projektu
Seznam ohlasů : 