Název:
Medial plate for treatment of distal tibia fractures
Citace
Theisz, G., Frydrýšek, K., Fojtík, F., Kubín, T., Pečenka, L., Demel, J., Madeja, R., Sadílek, M., Kratochvíl, J. a Pleva, L. Medial plate for treatment of distal tibia fractures.
In:
53rd Conference on Experimental Stress Analysis 2015-06-01 Český Krumlov.
Praha: Czech Technical University in Prague, 2015. s. 431-437. ISBN 9788001057346.
Podnázev
Rok vydání:
2015
Obor:
Chirurgie vč. transplantologie
Počet stran:
7
Strana od:
431
Strana do:
437
Forma vydání:
Tištená verze
Kód ISBN:
9788001057346
Kód ISSN:
Název sborníku:
53rd Conference on Experimental Stress Analysis
Sborník:
Mezinárodní
Název nakladatele:
Czech Technical University in Prague
Místo vydání:
Praha
Stát vydání:
Sborník vydaný v ČR
Název konference:
Místo konání konference:
Český Krumlov
Datum zahájení konference:
Typ akce podle státní
příslušnosti účastníků akce:
Evropská akce
Kód UT WoS:
EID:
Klíčová slova anglicky:
Biomechanics; Compression test; ESPI method; Failure; FEM; Medial plate; Orthopaedics; Strength analysis; Tibia; Traumatology
Popis v původním jazyce:
The objective of the present study was to carry out a strength and reliability assessment of a medial plate for the treatment of distal tibia fractures. This was performed via numerical modelling (FEM) and experimental methods (compression test and Electronic Speckle Pattern Interferometry - ESPI method). The plate is used for internal fixation in orthopaedics and traumatology. Analyses were performed for non-fused bone with comminuted fracture in the distal metaphysis (i.e. unsuccessful treatment, where all loads in the metaphysis are carried by the plate). An anatomical tibia model (based on CT images in Mimics software) was used for FE analysis. In the experiments, the bone was replaced with a shaped piece of spruce wood (based on Mimics software) and the plate was loaded until its failure (major plastic deformation). Numerical and experimental results were evaluated and compared.
Popis v anglickém jazyce:
The objective of the present study was to carry out a strength and reliability assessment of a medial plate for the treatment of distal tibia fractures. This was performed via numerical modelling (FEM) and experimental methods (compression test and Electronic Speckle Pattern Interferometry - ESPI method). The plate is used for internal fixation in orthopaedics and traumatology. Analyses were performed for non-fused bone with comminuted fracture in the distal metaphysis (i.e. unsuccessful treatment, where all loads in the metaphysis are carried by the plate). An anatomical tibia model (based on CT images in Mimics software) was used for FE analysis. In the experiments, the bone was replaced with a shaped piece of spruce wood (based on Mimics software) and the plate was loaded until its failure (major plastic deformation). Numerical and experimental results were evaluated and compared.